Volume 74, No.2: 2022 Siriraj Medical Journal
https://he02.tci-thaijo.org/index.php/sirirajmedj/index
93
Original Article
SMJ
In short, there were 4 subgroups in each study (for the
unopened eyedroppers and simulated use conditions)
as shown below:
(i) 0.01% atropine in HPMC at refrigerated temperature,
(ii) 0.01% atropine in HPMC at room temperature,
(iii) 0.01% atropine in BSS at refrigerated temperature,
(iv) 0.01% atropine in BSS at room temperature.
e eyedroppers stored at room temperature were
kept on the shelf, protected from light in their commercial
packages at 50%±10% residual humidity (RH).
Physicochemical and microbiological stability of
the 0.01% atropine ophthalmic solutions in simulated
use conditions
At day 0, 60 eyedroppers with two formulations
of 0.01% atropine solutions were prepared, with 15
eyedroppers for each subgroup (i–iv). For illustration
purposes, subgroup (i) with a total number of 15 eyedroppers
was used as an investigation process example. Each of the
15 eyedroppers was emitted daily (1 drop of the 0.01%
atropine solutions), that is, a drop was squeezed out
of the eyedropper and collected for analysis instead of
being dropped into the eye. Out of the 15 eyedroppers,
10 eyedroppers were obtained for visual inspection and
sterility assay. Next, 5 eyedroppers were tested at day 0,
15, and discarded. Another 5 eyedroppers were tested
at day 0 and 30. It is important to note the reason why
the eyedroppers were discarded aer the sterility assay
on day 15. Namely, subgroups (i) and (ii) both had an
approximate volume of 10 mL, while the sterility assay
required at least 4 mL. Hence, aer the daily emission and
two sterility assays, there would be an insucient amount
of solution remaining for another sterility assay and so
these were discarded. e remaining 5 eyedroppers from
the 15 totals were used for the atropine quantication
and pH measurements at days 0, 15, and 30.
Aer completing the 1-month study under simulated
use conditions, further investigations were planned, with the
aim to extend the experimental period of both formulations
at refrigerated temperature to 2 months. ere were 2
subgroups of eyedroppers here: (i) 6 eyedroppers of 0.01%
atropine in HPMC at refrigerated temperature, and (ii)
6 eyedroppers of 0.01% atropine in BSS at refrigerated
temperature. ese two subgroups were investigated
in the exact same manner as in the 1-month study. Out
of the 6 eyedroppers in each subgroup, 4 eyedroppers
were obtained for visual inspection and sterility assay at
day 0, one at another time point (days 15, 30, 45, or 60),
and one discarded (n = 1 for each time point/subgroup).
e remaining 2 from the 6 eyedroppers were used for
atropine quantication and pH measurements at days 0,
15, 30, 45, and 60 (n = 2 for each time point/subgroup).
Physicochemical and microbiological stability of
0.01% atropine ophthalmic solutions in the unopened
eyedroppers
In total, 120 eyedroppers of 0.01% atropine solutions
were prepared, comprising 30 eyedroppers for each
subgroup: atropine in HPMC at refrigerated temperature,
atropine in HPMC at room temperature, atropine in BSS
at refrigerated temperature, and atropine in BSS at room
temperature. In each subgroup, 5 unopened eyedroppers
were used for the analysis at days 30, 60, 90, 120, 150, and
180 (n = 5 for each time point/subgroup). Each eyedropper
was subjected to the following analyses: visual inspection,
atropine quantication, pH measurement, and sterility
assay. e baseline values for atropine quantication, pH
measurement, and the sterility assay were obtained from
the studies of the 0.01% atropine ophthalmic solutions
under the simulated use conditions.
Analyses
Quantication of atropine
The liquid chromatography with tandem mass
spectrometry (LC-MS/MS) method was applied for
quantitative analysis of the extemporaneously prepared
atropine solution. LC-MS/MS analysis was performed
using an Acquity Ultra Performance LC
TM
(Waters,
Co., Ltd. USA) coupled to a Quattro Premier XE Mass
Spectrometer (Micromass Technologies, UK) equipped
with an electrospray interface. For data acquisition and
processing, a MassLynx 4.1 SCN627 system (Micromass
Technologies, UK) was used.
Scopolamine hydrobromide was used as an internal
standard (IS). e chromatographic separation of atropine
and the internal standard was performed using a Kinetex
C18 column (50×2.10 mm, 1.7 µm; Phenomenex Ltd.,
USA). e mobile phase was an 85:15 (v/v) mixture of
0.1% (v/v) formic acid and acetonitrile in an isocratic
elution mode over a 2 min total run time. e ow rate
was 0.3 mL/min and the column temperature were set
at 30±5 °C. e injection volume was 1 μL. MS analyses
were carried out using the multiple reaction monitoring
(MRM) mode with positive electrospray ionization (ESI+).
e mass transition ion-pair was selected as m/z 290.1
to 124.1 for atropine and m/z 304.1 to 138.1 for the IS.
Validation of this method was performed according
to the International Conference on Harmonisation (ICH)
guidelines.
12
Linearity was determined by preparing
one calibration curve daily using six concentrations
of atropine (50, 100, 150, 200, 300, and 400 ng/mL),
obtained from atropine standard solution diluted in
diluent solutions (methanol and Milli-Q water at a ratio
of 1:1, v/v). e inuence of dierent weighting factors
(1/x and 1/x
2
) on the sum of the percentage relative error