Accuracy of the 256 Multi-detector Computerized Tomography in Detecting Coronary Artery Stenosis Experience from Bangkok Heart Hospital

Main Article Content

Lertlak Chaothawee, MD
Kittipan Visudharom, MD
Sucharath Warutama, MD
Pongpan Poonsawat, MD
Pradub Sukum, MD
Duddao Pantunurat
Yanipa Kaewlumhud
Darunee Jaiyong
Sangrawee Sri-ngam
Krittiya KhueanKhuab
Paweena Mapinta
Busakol Ngam-maung

Abstract

OBJECTIVE: To study the accuracy of the 256 Multi-detector Computerized Tomography (MDCT) in detecting coronary artery stenosis.


MATERIALS AND METHODS: We retrospectively analyzed angiographic findings of patients who underwent both 256 MDCT and invasive coronary angiography (ICA). All epicardial arteries, regardless of calcium burden, were segmented into proximal, mid and distal part for comparative analysis. Significant coronary artery stenosis was defined as the reduction of luminal diameter being equal to or more than 50%. The diagnostic accuracy of 256 MDCT in coronary artery stenosis evaluation was assessed by comparing its’ sensitivity, specificity, positive and negative predictive values to the gold standard ICA.


RESULTS: From January to December 2009, a total of 147 consecutive patients (124 male, 23 female, mean age of 60 ±12 years) underwent both MDCT and ICA were enrolled. Of total 1470 coronary segments (147 segments of LMA, 441 segments of the LAD, 441 segments of the LCx, 441 segments of the RCA), 98.9% were eligible to be assessed and only 1.1% (15/1470) were ineligible due to very high calcium clumps and severe motion artifacts. Compared to the ICA, the overall sensitivity of the 256 MDCT in detecting coronary stenosis was 88.3 %, specificity was 96.1%, positive predic- tive value was 88.1 % and the negative predictive value was 96.2% with an overall accuracy of 94.2% (p=0.20). In massive calcium scoring cases (calcium scoring ≥400 U), the sensitivity of 256 MDCT in detecting coronary artery stenosis was 90.3%, specificity was 96.1%, positive predictive value was 87.5% and the negative predictive value was 95.5%;  the  overall  accuracy  was  92.9%.  In nonmassive calcium scoring cases (calcium scoring.


CONCLUSION: The 256 MDCT, regardless of calcium burden, offers a reliable diagnostic accuracy in assessing coronary artery stenosis.


 

Article Details

How to Cite
1.
Chaothawee L, Visudharom K, Warutama S, Poonsawat P, Sukum P, Pantunurat D, Kaewlumhud Y, Jaiyong D, Sri-ngam S, KhueanKhuab K, Mapinta P, Ngam-maung B. Accuracy of the 256 Multi-detector Computerized Tomography in Detecting Coronary Artery Stenosis Experience from Bangkok Heart Hospital. BKK Med J [Internet]. 2011 Feb. 18 [cited 2024 Nov. 15];1(1):1. Available from: https://he02.tci-thaijo.org/index.php/bkkmedj/article/view/217521
Section
Original Article

References

1. Scanlon PJ, Faxon DP, Audet AM, Carabello B, Dehmer GJ, Eagle KA, Legako RD, Leon DF, Murray JA, Nissen SE, Pepine CJ, Watson RM, Ritchie JL, Gibbons RJ, Cheitlin MD, Eagle KA, Gardner TJ, Garson A, Russell RO Jr, Ryan TJ Jr, and Smith SC Jr. ACC/AHA guide lines for coronary angiography: A report of the Ameri- can College of ardiology/American Heart Association Task Force on Practice Guidelines (Committee on Coro- nary Angiography) developed in collaboration with the Society for Cardiac Angiography and Interventions J. Am. Coll. Cardiol 1999;33;1756-1824
2. Alderman EL, Stadius M. The angiographic definitions of the Bypass Angioplasty Revascularization Investiga- tion study (BARI). Coronary Artery Disease 1992;3: 1189-1207
3. Chagnaud S, Jacquier A, Paganelli F, Panuel M.Com- parison of 64 MDCT coronary CTA and coronary angio- graphy in the detection of coronary artery stenosis in low risk patients with stable angina and acute coronary syndrome. J Radiol 2009 Sep;90(9 Pt1):1055-1066
4. Leschka S, Alkadhi H, Plass A, Desbiolles L, Grunen- felder J, Marinek B, and Widermuth S.Accuracy of MSCT coronary angiography with 64 sslice technology : first experience . Eur Heart J 2005;26:1482-1487
5. Raff GL, Gallagher MJ, O’Neill WW, Goldstein JA. Diagnostic Accuracy of Noninvasive Coronary Angiog- raphy Using 64-Slice Spiral Computed Tomography, J Am Coll Cardiol 2005;46:552–557
6. Joseph A Abro; U.Joseph Schoepf ; Walter Huda. Effec- tive patient Radiation doses and skin organ doses at coro- nary CT angiography and invasive coronary angiography. Circulation 2008:118:598-599
7. Hsiao EM, Rybicki FJ, and Steigner M. CT Coronary Angiography: 256-Sliceand 320-Detector Row Scanners. Curr Cardiol Rep 2010 January ; 12(1):68-75