Characterization of Candidate Chemical Reference Substances for In Vitro Cytotoxicity Testing (ISO 10993-5) Using the L929 Mouse Fibroblast Cell Line

Authors

  • Luxkana Truatnok Laboratory Analysis Unit, National Laboratory Animal Center, Mahidol University, Nakhon Pathom, 73170, Thailand
  • Nathapong Sukhawipat NAS Laboratory, Department of Mechanical Engineering Technology, King Mongkut’s University of Technology North Bangkok, Bangkok, 10800, Thailand

Keywords:

Cytotoxicity, ISO 10993-5, Positive Control, Negative Control, L929 Cells, MTT Assay

Abstract

Reliable reference materials and appropriate working concentrations are essential for ensuring the performance of in vitro cytotoxicity assays conducted according to ISO 10993-5. Although the standard recommends reference materials for assay verification, it does not specify working concentrations for soluble control substances that consistently produce defined cytotoxic responses in L929 cells. This study characterized the dose–response profiles of five reference substances, including zinc acetate, zinc chloride, Triton X-100, zinc diethyldithiocarbamate (ZDEC), and high-density polyethylene (HDPE), to identify laboratory-specific working concentrations suitable for routine in vitro cytotoxicity testing. L929 mouse fibroblasts were exposed to serial concentrations of the test substances, and cell viability was determined using the MTT assay following ISO 10993-5. All soluble control substances produced concentration-dependent reductions in cell viability, with ZDEC exhibiting the greatest cytotoxic potency, followed by Triton X-100, zinc chloride, and zinc acetate. HDPE extract also exhibited concentration-dependent effects; however, only the 5% extract maintained cell viability above the 70% cell viability threshold used for cytotoxicity classification, whereas higher extract concentrations reduced cell viability below this threshold. Based on the observed dose–response relationships, candidate laboratory-specific working concentrations were identified for each candidate control substance under the experimental conditions employed in this study. These findings provide a practical basis for selecting positive and negative control concentrations for routine ISO 10993-5 cytotoxicity testing using L929 cells under comparable laboratory conditions but should not be interpreted as constituting a formal validation of the assay method.

References

Bae SH, Che JH, Seo JM, Jeong J, Kim ET, Lee SW, et al. In vitro biocompatibility of various polymer-based microelectrode arrays for retinal prosthesis. Invest Ophthalmol Vis Sci. 2012;53(6):2653-7.

Borenfreund E, Puerner JA. Toxicity determined in vitro by morphological alterations and neutral red absorption. Toxicol Lett. 1985;24(2-3):119-24.

Fotakis G, Timbrell JA. In vitro cytotoxicity assays: comparison of LDH, neutral red, MTT and protein assay in hepatoma cell lines following exposure to cadmium chloride. Toxicol Lett. 2006;160(2):171-7.

Gruber S, Nickel A. Toxic or not toxic? the specifications of the standard ISO 10993-5 are not explicit enough to yield comparable results in the cytotoxicity assessment of an identical medical device. Front Med Technol. 2023;5:1195529.

International Organization for Standardization. ISO 10993-5:2009 Biological evaluation of medical devices—Part 5: Tests for in vitro cytotoxicity [monograph on the Internet]. Geneva: International Organization for Standardization; 2009 [cited 2026 Jan 23]. Available from: https://www.iso.org/standard/36406.html.

International Organization for Standardization. ISO 10993-12:2021 Biological evaluation of medical devices—Part 12: Sample preparation and reference materials [monograph on the Internet]. Geneva: International Organization for Standardization; 2021 [cited 2026 Jan 23]. Available from: https://www.iso.org/standard/75691.html.

Li W, Zhou J, Xu Y. Study of the in vitro cytotoxicity testing of medical devices. Biomed Rep. 2015;3(5):617-20.

Mosmann T. Rapid colorimetric assay for cellular growth and survival: application to proliferation and cytotoxicity assays. J Immunol Methods. 1983;65(1-2):55-63.

Park JC, Park BJ, Lee DH, Suh H, Kim DG, Kwon OH. Evaluation of the cytotoxicity of polyetherurethane (PU) film containing zinc diethyldithiocarbamate (ZDEC) on various cell lines. Yonsei Med J. 2002;43(4):518-26.

Riss TL, Moravec RA, Niles AL, Duellman S, Benink HA, Worzella TJ, et al. Cell viability assays. In: Markossian S, Grossman A, Baskir H, et al., editors. Assay Guidance Manual. Bethesda (MD): Eli Lilly & Company and the National Center for Advancing Translational Sciences; 2016. p. 1-25.

Tua-ngam P, Jira-anankul N, Anuwongnukroh N, Dechkunakorn S, Laokijcharoen P. Cytotoxicity of three commercial orthodontic elastomeric ligature brands. Matec Web Conf. 2017;108:06004.

Van Tienhoven EA, Korbee D, Schipper L, Verharen HW, De Jong WH. In vitro and in vivo cytotoxicity assays using PVC and LDPE as model materials. J Biomed Mater Res A. 2006;78(1):175-82.

Wang MO, Etheridge JM, Thompson JA, Vorwald CE, Dean D, Fisher JP. Evaluation of the in vitro cytotoxicity of cross-linked biomaterials. Biomacromolecules. 2013;14(5):1321-9.

Weyermann J, Lochmann D, Zimmer A. A practical note on the use of cytotoxicity assays. Int J Pharm. 2005;288(2):369-76.

Wypych G. Handbook of Polymers. 2nd ed. Toronto: ChemTec Publishing; 2018.

Downloads

Published

2026-11-03

How to Cite

Truatnok, L., & Sukhawipat, N. (2026). Characterization of Candidate Chemical Reference Substances for In Vitro Cytotoxicity Testing (ISO 10993-5) Using the L929 Mouse Fibroblast Cell Line. Journal of Applied Animal Science, 20(1), 25–37. Retrieved from https://he02.tci-thaijo.org/index.php/jaas_muvs/article/view/282252

Issue

Section

Research Articles