Assessing Agreement of Intraocular Tonometry Between Non Contact and Goldman Applanation Tonometer
Keywords:
IOP, noncontact tonometry, Goldman applanation tonometry, agreementAbstract
Objectives: To study intraocular pressure measured by non contact tonometry compare with intraocular pressure measured by Goldman applanation tonometry.
Material and methods: The intraocular pressure of out-patients in division of ophthalmology Nakhonpathom hospital was measured by non contact tonometer by trained medical personnels and then repeated it on the same eye with Goldman applanation tonometer by studying ophthalmologist. The data of intraocular pressure from both methods were compared by using unpaired t-test and analyzed to reach the agreement with linear regression analysis and Bland Altman test
Results: Intraocular pressure of 248 eyes were measured with both non contact tonometer and Goldman applanation tonometer. There was no statistic difference in mean pressure (p-value = 0.3065), correlation coefficient r = 0.908, slope(t) = 1.0236 and the mean difference of Bland Altman test = 0.5 mmHg (2 S.D. = 4.5 mmHg)
Conclusion: Non contact tonometry was comparative to Goldman applanation tonometry and there was statistic agreement between two methods.
Downloads
Published
How to Cite
Issue
Section
License
ลิขสิทธิ์บทความเป็นของผู้เขียนบทความ แต่หากผลงานของท่านได้รับการพิจารณาตีพิมพ์ลงวารสารแพทย์เขต 4-5 จะคงไว้ซึ่งสิทธิ์ในการตีพิมพ์ครั้งแรกด้วยเหตุที่บทความจะปรากฎในวารสารที่เข้าถึงได้ จึงอนุญาตให้นำบทความในวารสารไปใช้ประโยชน์ได้ในเชิงวิชาการโดยจำเป็นต้องมีการอ้างอิงถึงชื่อวารสารอย่างถูกต้อง แต่ไม่อนุญาตให้นำไปใช้ในเชิงพาณิชย์