Comparison of Hemoglobin Levels between Point-of-Care Testing and Automated Hematology Analyzer in Patients at Samutsakhon Hospital
Keywords:
Hemoglobin, Point of care testing (POCT), Automated hematology analyzer, Comparative analysis, Test result correlationAbstract
Objective: To compare hemoglobin (Hb) measurement results between a point-of-care testing (POCT) device and an automated cell counter in a central laboratory.
Methods: This analytical study included 50 participants whose Hb levels were measured using both a POCT device and an automated cell counter. The results were compared using mean difference analysis, Pearson correlation, and Bland-Altman plots.
Results: The mean Hb level measured by the POCT device was 12.0 ± 2.5 g/dl, while that measured by the automated cell counter was 12.7 ± 2.6 g/dl, with no statistically significant difference (p > .05). Pearson correlation analysis demonstrated a very high positive correlation (r = 0.993; p < .001). Bland-Altman analysis showed a mean difference of -0.71 g/dl (95% CI: -0.80 to -0.62; p < .001), indicating a slight negative bias from the POCT device.
Conclusion: The POCT device demonstrated high agreement with the automated cell counter and can be considered a reliable alternative for Hb measurement, particularly in settings where rapid results are required.
References
คมกฤษ ศรีสรรพศิริกุล. Erythrocyte abnormalities. ใน: พรรธนมณฑน์ อุชชิน, ชุติธร เกตุลอย, อรุณี ทองอัครนิโรจน์, รชต วรเวชวัฒนะ, บรรณาธิการ. พยาธิวิทยาคลินิก Clinical pathology. กรุงเทพมหานคร: จุฬาลงกรณ์มหาวิทยาลัย; 2555. 348–55.
Silver RT, Chow W, Orazi A, Arles SP, Goldsmith SJ. Evaluation of who criteria for diagnosis
of polycythemia vera: A prospective analysis. Blood. 2013;122(11):1881–6.
doi:10.1182/blood-2013-06-508416
ศูนย์รับบริจาคโลหิตแห่งชาติ สภากาชาดไทย. คู่มือการคัดเลือกผู้บริจาคโลหิต. พิมพ์ครั้งที่ 6.สมุทรปราการ: อุดมศึกษา; 2560.
Chutipongtanate A, Yasaeng C, Virankabutra T, Chutipongtanate S. Systematic comparison
of four point-of-care methods versus the reference laboratory measurement of
hemoglobin in the surgical ICU setting: A cross-sectional method comparison study.
BMC Anesthesiology. 2020;20(1):92. doi:10.1186/s12871-020-01008-8
นภชาญ เอื้อประเสริฐ. Analysis of blood cells by automation. ใน: นภชาญ เอื้อประเสริฐ, จันทนา ผลประเสริฐ, บรรณาธิการ. Diagnostic approach in hematology: Laboratory interpretation and clinical correlation. 2nd ed. กรุงเทพมหานคร: จุฬาลงกรณ์มหาวิทยาลัย; 2560. 17.
พรรธนมณฑน์ อุชชิน. Point-of-care testing. ใน: พรรธนมณฑน์ อุชชิน, ชุติธร เกตุลอย, อรุณี ทองอัครนิโรจน์, รชต วรเวชวัฒนะ, บรรณาธิการ. พยาธิวิทยาคลินิก Clinical pathology. กรุงเทพมหานคร: จุฬาลงกรณ์มหาวิทยาลัย; 2555. 114.
Chantkran W, Jamnarnwej P, Sritanabutr P, Arnutti P. Evaluation of point-of-care testing
device for anemia detection: A cross-sectional method comparison study from
Thailand. J Clin Lab Anal. 2021;35(10): e23976 doi:10.1002/jcla.23976
Arhamidolatabadi A, Hatamabadi H, Amini K. Point-of-care hemoglobin measurement in
comparison with hematology analyzer: A cross-sectional study in emergency department. Front Emerg Med. 2019;4(4):e82.
Herman J, Park B, Awsare B, West F, Crittendon D, Evans L, et al. Point-of-care versus central testing of hemoglobin during large volume blood transfusion. BMC Anesthesiology.
;19(1):240. doi:10.1186/s12871-019-0916-2
Grech K, Zammit V. Point-of-care testing effectiveness on blood donor hemoglobin
testing. Lab Med. 2021;52(4):364–8. doi:10.1093/labmed/lmaa102
Dolscheid-Pommerich RC, Dolscheid S, Grigutsch D, Stoffel-Wagner B, Graeff I. Comparability of point-of-care versus central laboratory hemoglobin determination in emergency patients at a Supra-maximal care Hospital. PLOS ONE. 2016;11(11): e0166521. doi:10.1371/journal.pone.0166521
Osborn ZT, Villalba N, Derickson PR, Sewatsky TP, Wager AP, Freeman K. Accuracy of
point-of-care testing for anemia in the emergency department. Respir Care. 2019;64(11):1343–50. doi:10.4187/respcare.06364
Killilea DW, Kuypers FA, Larkin SK, Schultz K. Blood draw site and analytic device
influence hemoglobin measurements. PLOS ONE. 2022;17(11):e0278350. doi:10.1371/journal.pone.0278350
Ramaswamy G, Jaiswal A, Vohra K, Kaur R, Bairwa M, Singh A, et al. Correction equation
for hemoglobin values obtained using point of care tests—a step towards realistic anemia burden estimates. Diagnostics (Basel). 2022;12(12):3191. doi:10.3390/diagnostics12123191
Ahn HS, Lenet T, Gilbert RWD, Mallick R, Shaw JL, Fergusson DA, et al. Accuracy of
point-of-care testing devices for haemoglobin in the operating room: Meta-analysis.
BJS Open. 2024;8(1):zrad148. doi:10.1093/bjsopen/zrad148
Shah N, Osea EA, Martinez GJ. Accuracy of noninvasive hemoglobin and invasive point-
of-care hemoglobin testing compared with a laboratory analyzer. Int J Lab Hematol. 2014;36(1):56–61. doi:10.1111/ijlh.12118
Downloads
Published
How to Cite
Issue
Section
License
Copyright (c) 2026 Region 4-5 Medical Journal

This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.
ลิขสิทธิ์บทความเป็นของผู้เขียนบทความ แต่หากผลงานของท่านได้รับการพิจารณาตีพิมพ์ลงวารสารแพทย์เขต 4-5 จะคงไว้ซึ่งสิทธิ์ในการตีพิมพ์ครั้งแรกด้วยเหตุที่บทความจะปรากฎในวารสารที่เข้าถึงได้ จึงอนุญาตให้นำบทความในวารสารไปใช้ประโยชน์ได้ในเชิงวิชาการโดยจำเป็นต้องมีการอ้างอิงถึงชื่อวารสารอย่างถูกต้อง แต่ไม่อนุญาตให้นำไปใช้ในเชิงพาณิชย์