The Value of Local ISI Calibration in Correcting the Variability in INR Determination

Authors

  • Wanida Wongtiraporn Department of Clinical Pathology, Faculty of Medicine Siriraj Hospital, Mahidol University
  • Nisarat Opartkiattikul Department of Clinical Pathology, Faculty of Medicine Siriraj Hospital, Mahidol University
  • Panutsaya Tientadakul Department of Clinical Pathology, Faculty of Medicine Siriraj Hospital, Mahidol University

Keywords:

-

Abstract

We have compared the International Normalized Radio (INR) determination using the manufactures' stated International Sensitivity Index (ISI) with an alternative method using local calibration of ISI with the calibrator plasma. It was found that the variability of the INR was less when the results were expressed by INR using local calibration of ISI. The results indicated that the local ISI calibration might reduce the variability in the INR determination.

Downloads

Published

01-07-2003

How to Cite

Wongtiraporn, W. ., Opartkiattikul, N. ., & Tientadakul, P. . (2003). The Value of Local ISI Calibration in Correcting the Variability in INR Determination. Siriraj Medical Journal, 55(7), 381–384. Retrieved from https://he02.tci-thaijo.org/index.php/sirirajmedj/article/view/245416

Issue

Section

Original Article