Radiation Measurement of Handheld Dental X-ray Devices in Central Region of Thailand

Authors

  • Anong Singkavongsay Bureau of Radiation and Medical Devices, Department of Medical Sciences
  • Suphawat Thupsuri Bureau of Radiation and Medical Devices, Department of Medical Sciences

Keywords:

Radiation output, Scatter radiation, Handheld dental x-ray device

Abstract

       This study was to compare radiation output and scatter radiation between handheld and fixed dental X-ray devices used at diagnostic radiology laboratories of hospitals in central region, Thailand during 2016-2018. A total of 300 dental X-ray devices (150 each handheld and fixed units) from laboratories in Bangkok, Nonthaburi, Pathumthani, Ayutthaya, Singburi, Angthong, and Saraburi were evaluated. Dose meter and Survey meter were used for measuring radiation output and scatter radiation, respectively. Scatter radiation was measured at 5 different positions; at handles, at 1-, 2-, 3- meters from irradiating units and at staff area. It was found that an average radiation output of handheld devices was 1,135 μGy while the average of scatter radiation at the aforementioned positions was 1,130, 66, 10, 2 and 125 μSv/h, respectively.
As for fixed dental X-ray devices, the average radiation output was 2,711 μGy whereas the average of scatter radiation at 5 different positions was 1,885, 149, 19, 4 and 0 μSv/h, respectively. This indicated that operators tend to have more radiation exposure using the handheld dental X-ray devices. Radiation safety should be concerned for both operators and customers.

References

สำนักงานปรมาณูเพื่อสันติ. การป้องกันอันตรายจากรังสี. กรุงเทพฯ: โรงพิมพ์คุรุสภา; 2546.

European guidelines on radiation protection in dental radiology: the safe use of radiographs in dental practice. Issue No.136. Luxembourg: European Communities; 2004. p. 41-49.

Conference of Radiation Control Program Directors (CRCPD). QA collectible: hand-held dental x-ray units. [online]. 2010. Aug [cited 2019 Mar 30]; [6 screens]. Available from: URL: https://www.crcpd.org/resource/resmgr/docs/QA_Collectibles/Hand-heldDentalX-rayUnitsQAC.pdf.

U.S. Food and Drug Administration. Illegal sale of potentially unsafe hand-held dental x-ray units: FDA safety communication. [online]. 2012; [cited 2019 Mar 30]; Available from: URL: https://www.fda.gov/Radiation-EmittingProducts/RadiationSafety/AlertsandNotices/ucm291214.htm.

Gray JE, Bailey ED, Ludlow JB. Dental staff doses with handheld dental intraoral x-ray units. Health Phys 2012; 102(2): 137-42.

Kim EK. Effect of amount of battery charge on tube voltage in different hand-held dental x-ray systems. Imaging Sci Dent 2012; 42: 1-4.

Lee MJ, Seo JH, Lee MG, Choi YH. Leakage and scattered radiation dosage in portable dental x-rays. Int J Clin Prev Dent 2013; 9(3): 131-8.

Mahdian M, Pakchoian AJ, Dagdeviren D, Alzahrani A, Jalali E, Tadinada A, et al. Using hand-held dental x-ray devices: ensuring safety for patients and operators. JADA 2014; 145(11): 1130-2.

ศิริวรรณ จูเลียง, สายัณห์ เมืองสว่าง. ความปลอดภัยจากการใช้เครื่องเอกซเรย์ฟันในเขตสาธารณสุขที่ 7. ว กรมวิทย พ 2556; 55(4): 236-45.

กรมวิทยาศาสตร์การแพทย์ กระทรวงสาธารณสุข. ค่าปริมาณรังสีอ้างอิงจากการถ่ายรังสีวินิจฉัยด้วยเครื่องเอกซเรย์ฟันชนิดถ่ายภายในช่องปาก. [ออนไลน์]. 2561; [สืบค้น 30 เมษายน 2562]; [10 หน้า]. เข้าถึงได้จาก: URL: https://dmscsmartlifeblog.com/userfiles/files/ค่าปริมาณรังสีอ้างอิงDent-final(1).pdf.

กรมวิทยาศาสตร์การแพทย์ กระทรวงสาธารณสุข. มาตรฐานคุณภาพเครื่องเอกซเรย์วินิจฉัยกรมวิทยาศาสตร์การแพทย์ พ.ศ. 2558. กรุงเทพฯ: โรงพิมพ์ชุมนุมสหกรณ์การเกษตรแห่งประเทศไทย; 2558.

Makdissi J, Pawar RR, Johnson B, Chong BS. The effects of device position on the operator’s radiation dose when using a handheld portable x-ray device. Dentomaxillofac Radiol 2016; 45(3): 20150245. (7 p.).

Downloads

Published

30-09-2019

How to Cite

1.
Singkavongsay A, Thupsuri S. Radiation Measurement of Handheld Dental X-ray Devices in Central Region of Thailand. ว กรมวิทย พ [internet]. 2019 Sep. 30 [cited 2026 Mar. 4];61(3):96-106. available from: https://he02.tci-thaijo.org/index.php/dmsc/article/view/240692

Issue

Section

General Articles