Quality of Non-invasive Sphygmomanometers in District Health Promoting Hospitals of the 4th Regional Health Promoting Center
Keywords:
non-invasive sphygmomanometers, maximum permissible errors, air leakageAbstract
Quality of non-invasive sphygmomanometers in district health promoting hospitals of the 4th regional health promoting center was evaluated. Two thousand and seventy-two machines including 335 of mechanical type and 1,737 of automated type were evaluated during the iscal years of 2013 - 2016. According to the international standard-based Testing Procedures for Medical Measuring Devices TP-MMD-01: Blood Pressure Monitor issued by the Ministry of Public Health of Thailand, the maximum permissible errors (MPE) of the cuf pressure were tested to investigate measurement accuracy of the machines. In addition, air leakage testing rate was measured to investigate integrity of the machine, tube and cuf. For mechanical sphygmomanometers, 21 machines (6.3%) were reported to be failed to meet the MPE criteria due to air bubbles in the tube, possibly contaminated from mercury or leakage of mercury. Sixty-two machines (18.5%) failed to meet the air leakage criteria due to defectiveness of tube, bulb or cuf. Furthermore, 5 machines (1.5%) failed to meet both MPE and air leakage criteria. For automated sphygmomanometers, 28 machines (1.6%) were determined to be failed to meet the MPE criteria because reading values were deviated from reference values of greater than 4 mmHg. Two hundred and forty-two machines (13.9%) failed to meet air leakage criteria because of the defectiveness of air tube or cuf. Moreover, 5 machines (0.5%) failed to meet both MPE and air leakage criteria. Root-cause analysis should be determined after quality of sphygmomanometers was poorly detected. Change or repair of any parts of the machines will result in not only reliability of the machines but also accuracy on diagnosis and monitoring of diseases.
References
สำนักโรคไม่ติดต่อ กรมควบคุมโรค กระทรวงสาธารณสุข. รายงานประจำปี 2559. กรุงเทพฯ : สำนักงานกิจการโรงพิมพ์องค์การสงเคราะห์ทหารผ่านศึก ในพระบรมราชูปถัมภ์; 2559.
กระทรวงสาธารณสุข. คู่มือการทดสอบเครื่องมือวัดทางการแพทย์ TP-MMD-01: เครื่องวัดความดันโลหิต. กรุงเทพฯ : บริษัท บียอนด์ พับลิสซิ่ง จำกัด; 2559.
OIML R 16-1:2002. Non-invasive mechanical sphygmomanometers. France: Organization Internationale de Metrologie Legale; 2002.
ISO 81060-1:2007. Non-invasive sphygmomanometers -- Part 1: requirements and test methods for non-automated measurement type. Geneva: International Organization for Standardization; 2007.
OIML R 16-2:2002. Non-invasive automated sphygmomanometers. France: Organization Internationale de Metrologie Legale; 2002.
IEC 80601-2-30:2009. Medical electrical equipment --Part 2-30: particular requirements for basic safety and essential performance of automated non-invasive sphygmomanometers. Geneva: International Organization for Standardization; 2009.
รุ้งระวี นาวีเจริญ. ความดันโลหิตสูง เทคนิคควบคุมความดันด้วยตัวเองอย่างได้ผล. กรุงเทพฯ : สำนักพิมพ์อมรินทร์สุขภาพ; 2552.
เอกสารการประชุมสัมมนาเรื่อง Seminar on the testing of non-invasive sphygmomanometers. วันที่ 2-6 สิงหาคม 2553. กรุงเทพฯ : สถาบันมาตรวิทยาแห่งชาติ กระทรวงวิทยาศาสตร์และเทคโนโลยี; 2553.
กรมวิทยาศาสตร์การแพทย์ กระทรวงสาธารณสุข. โรคความดันโลหิตสูงและเครื่องวัดความดันโลหิต. กรุงเทพฯ : บริษัท 1241 มิราคูลัส จำกัด; 2555.
Downloads
Published
How to Cite
Issue
Section
License

This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.