Radiation shielding evaluation of the X-ray room in small private clinics

Authors

  • Khatayut Nigapruke Regional Medical Sciences Center 8 Udon Thani, Department of Medical Science

Keywords:

x-ray room, dose limit

Abstract

       The purpose of this study was to explore and evaluate the safety of the X-ray room from scattered and leakage radiation doses arising from diagnostic and dental radiography for 13 small private clinics that have not passed safety testing. The digital survey meter was used for measurement of the radiation dose at different locations both inside and outside the X-ray rooms as well as the adjacent neighbors. The results showed that occupational dose, public dose for person who lives in the same building, and public dose for person who lives in the adjacent building or neighbor were 0.5 - 50.1 μGy/wk, 2.9 - 95.4 μGy/wk, and 0.2 - 3.1 μGy/wk, respectively. It was found that members of the public group who live in the same building from 10 small private clinics (77%) received the radiation dose exceed 20 uGy/wk, which is the dose limit recommended by the International Commission on Radiological Protection (ICRP) report No. 103. In addition, the calculation of barrier or wall thickness was made according to the recommendation of the National Council on Radiation Protection and Measurements (NCRP) report no.145 and no. 147, and should be served as a guideline of the x-ray room for small private clinics in future.

References

The 2007 Recommendations of the International Commission on Radiological Protection. ICRP Publication 103. Ann. ICRP 37 (2-4).

NCRP Report No. 147. Structural shielding design for medical X-ray imaging facilities. Bethesda, MD: National Council on Radiation Protection and measurements; 2004.

NCRP Report No. 145. Radiation protection in dentistry. Bethesda, MD: National Council on Radiation Protection and Measurements; 2003.

Simpkin DJ. Diagnostic X-ray shielding calculations for effective dose equivalent (abstract). Med Phys 1994; 21: 893.

Simpkin DJ. XRAYBARR, Documentation. Mukwonago, Wisconsin 53149; 1996.

Pesianian I, Mesbahi A, Shafaee A. Shielding evaluation of a typical radiography department: a comparison between NCRP reports No.49 and 147. Iran J Radiat Res 2009; 6 (4): 183-188.

Radiation shielding requirements for diagnostic x-ray facilities. Western Australia: Radiation Health Branch; 2004.