Determination of A Correction Factor for Measurement of Occupational Dose from Beta Source Using InLight®nanoDot OSL

Authors

  • Praphat Lapcharoenkit Bureau of Radiation and Medical Devices, Department of Medical Sciences

Keywords:

InLight®nanoDot OSL, personal dosimeter

Abstract

       An occupational radiation worker must have a personal dosimeter to monitor occupational exposure dose. InLight®nanoDot OSL, is the dosimeter used for detection of exposure at a specific position of the body. Accurate determination of occupational dose of exposure from beta sources needs a correction factor for energy. This study was conducted to obtain the correction factor essential for InLight®nanoDot OSL after exposure to Sr-90, the beta source, at 0, 5, 10, 15 and 20 millisievert (mSv). The correction factor was shown to be 2.90 with differences ranging from 0.20% to 2.93%. Using the correction obtained, the amount of beta radiation measured was closed to that radiated from the Sr-90 source. Our study suggested that determination of the correction factor is necessary prior to performing a service on measurement of occupational dose from beta sources using InLight®nanoDot OSL.

References

Botter-Jensen L, McKeever SWS, Wintle AG. Optically stimulated luminescence dosimetry. Amsterdam: Elsevier Science B.V.; 2003.

Landauer. Personal dosimetry performance testing for InLight dosimetry conducted at Pacific Northwest National Laboratory. Glenwood, IL: Landauer; 2006

Whole Body Dose, Algorithm for the Landauer. InLight basic-OSLN dosimeter, Revision: InLight_Basic_OSNL_

N_2008 09/19/08. Gaithersburg, MD: Validated by National Voluntary Laboratory Accreditation Program (NVLAP); 2008.

International Atomic Energy Agency. Calibration of radiation protection monitoring instruments. Vienna: IAEA; 2000.

Sudchai W, Thierrattanakul S, Jianklang S, Termsuk C. Nanodot dosimeter for extremity monitoring service in Thailand. Progress in 4th Asian and Oceanic Congress for Radiation Protection; 2014 May 12-16. Kuala Lumpur,

Malaysia: AOCRP; 2014.

International Atomic Energy Agency. Assessment of occupational exposure due to external source of radiation: safety guide. Vienna: IAEA; 1999.

Downloads

Published

29-06-2018

How to Cite

1.
Lapcharoenkit P. Determination of A Correction Factor for Measurement of Occupational Dose from Beta Source Using InLight®nanoDot OSL. ว กรมวิทย พ [internet]. 2018 Jun. 29 [cited 2026 Sep. 29];60(2):76-84. available from: https://he02.tci-thaijo.org/index.php/dmsc/article/view/241083