Comparison of X-ray Beam Overlap Measurements between Digital X-ray Ruler and Collimator Test Tool

Comparison of X-ray Beam Overlap Measurements

Authors

  • Rungtip Putburee Regional Medical Sciences Center 8 Udonthani
  • Khatayut Nigapruke Regional Medical Sciences Center 9 Nakhon Ratchasima

Keywords:

Digital X-ray ruler, Collimator test tool, X-ray beam overlap measurements

Abstract

         A newly developed tool, the Digital X-ray ruler, has been designed for evaluating X-ray beam and beam overlap. The objective of this study was to compare X-ray beam overlap measurements between the Digital X-ray ruler and the collimator test tool, which was the current standard device used for measuring beam overlap in general radiography X-ray machine. The test results comparing X-ray beam overlap with the light field revealed that in clear-beam general diagnostic X-ray machines, the digital X-ray ruler had a maximum difference from the collimator test tool at 0.03 centimeters, accounting for 0.34%. Meanwhile, in general diagnostic X-ray machines with unclear beams, the maximum difference between the beam overlap and the light field was 0.09 centimeters, calculated as 1.41%. Therefore, it could be summarized that there was no significant difference in measurement results between the two instruments. The Digital X-ray ruler can be used as a suitable alternative to the collimator test tool.

 

References

UNSCEAR. Sources and effects of ionizingradiation volume 1 & 2: UNSCEAR 2008 report to the General Assembly with scientificannexes. New York: United Nations Publication; 2008.

Willis CE, Slovis TL. The ALARA concept in pediatric CR and DR: dose reduction in pediatric radiographic exams--a white paper conference executive summary. Pediatr Radiol 2004; 34(Suppl 3): S162-4.

Cook JV, Kyriou JC, Pettet A, Fitzgerald MC, Shah K, Pablot SM. Key factors in the optimization of paediatric X-ray practice. Br J Radiol 2001; 74(887): 1032-40.

Gallet JM, Reed MH, Hlady J. A novel quality assurance method in a university teaching paediatric radiology department. Br J Radiol 2000; 73(872): 843-6.

ICRP. Recommendations of the International Commission on Radiological Protection. Ottawa: ICRP Publication 60 Ann ICRP 21: 1-3; 1991.

Faculty of Clinical Radiology. General X-ray QA and QC guideline. Sydney, Australia: The Royal Australian and New Zealand College of Radiologists; 2013.

กรมวิทยาศาสตร์การแพทย์ กระทรวงสาธารณสุข. มาตรฐานคุณภาพเครื่องเอกซเรย์วินิจฉัยทางการแพทย์ (Quality standard of medical diagnostic X-ray machines). กรุงเทพฯ: บริษัท บียอนด์ พับลิสชิ่ง จำกัด; 2562. หน้า 3-19.

Stollfuss J, Schneider K, Krüger-Stollfuss I. A comparative study of collimation in bedside chest radiography for preterm infants in two teaching hospitals. Eur J Radiol Open 2015; 2: 118-22.

QUART Nonius. User guide electronic X-ray ruler for radiation field edges and fan-beam evaluation, version 07/18. Kirchenweg: Quart GmbH; 2019.

Vashista M, Paul S. Correlation between full width at half maximum (FWHM) of XRD peak with residual stress on ground surfaces. Philosophical Magazine 2012; 92(33): 4194-204.

International Atomic Energy Agency. Quality assurance programme for digital mammography. IAEA human health series no. 17. Vienna: IAEA; 2011. p. 3-24

Downloads

Published

26-12-2023

How to Cite

1.
Putburee R, Nigapruke K. Comparison of X-ray Beam Overlap Measurements between Digital X-ray Ruler and Collimator Test Tool: Comparison of X-ray Beam Overlap Measurements. ว กรมวิทย พ [internet]. 2023 Dec. 26 [cited 2025 Dec. 13];65(4):248-55. available from: https://he02.tci-thaijo.org/index.php/dmsc/article/view/261125

Issue

Section

Original Articles