Comparison of X-ray Beam Overlap Measurements between Digital X-ray Ruler and Collimator Test Tool
Comparison of X-ray Beam Overlap Measurements
Keywords:
Digital X-ray ruler, Collimator test tool, X-ray beam overlap measurementsAbstract
A newly developed tool, the Digital X-ray ruler, has been designed for evaluating X-ray beam and beam overlap. The objective of this study was to compare X-ray beam overlap measurements between the Digital X-ray ruler and the collimator test tool, which was the current standard device used for measuring beam overlap in general radiography X-ray machine. The test results comparing X-ray beam overlap with the light field revealed that in clear-beam general diagnostic X-ray machines, the digital X-ray ruler had a maximum difference from the collimator test tool at 0.03 centimeters, accounting for 0.34%. Meanwhile, in general diagnostic X-ray machines with unclear beams, the maximum difference between the beam overlap and the light field was 0.09 centimeters, calculated as 1.41%. Therefore, it could be summarized that there was no significant difference in measurement results between the two instruments. The Digital X-ray ruler can be used as a suitable alternative to the collimator test tool.
References
UNSCEAR. Sources and effects of ionizingradiation volume 1 & 2: UNSCEAR 2008 report to the General Assembly with scientificannexes. New York: United Nations Publication; 2008.
Willis CE, Slovis TL. The ALARA concept in pediatric CR and DR: dose reduction in pediatric radiographic exams--a white paper conference executive summary. Pediatr Radiol 2004; 34(Suppl 3): S162-4.
Cook JV, Kyriou JC, Pettet A, Fitzgerald MC, Shah K, Pablot SM. Key factors in the optimization of paediatric X-ray practice. Br J Radiol 2001; 74(887): 1032-40.
Gallet JM, Reed MH, Hlady J. A novel quality assurance method in a university teaching paediatric radiology department. Br J Radiol 2000; 73(872): 843-6.
ICRP. Recommendations of the International Commission on Radiological Protection. Ottawa: ICRP Publication 60 Ann ICRP 21: 1-3; 1991.
Faculty of Clinical Radiology. General X-ray QA and QC guideline. Sydney, Australia: The Royal Australian and New Zealand College of Radiologists; 2013.
กรมวิทยาศาสตร์การแพทย์ กระทรวงสาธารณสุข. มาตรฐานคุณภาพเครื่องเอกซเรย์วินิจฉัยทางการแพทย์ (Quality standard of medical diagnostic X-ray machines). กรุงเทพฯ: บริษัท บียอนด์ พับลิสชิ่ง จำกัด; 2562. หน้า 3-19.
Stollfuss J, Schneider K, Krüger-Stollfuss I. A comparative study of collimation in bedside chest radiography for preterm infants in two teaching hospitals. Eur J Radiol Open 2015; 2: 118-22.
QUART Nonius. User guide electronic X-ray ruler for radiation field edges and fan-beam evaluation, version 07/18. Kirchenweg: Quart GmbH; 2019.
Vashista M, Paul S. Correlation between full width at half maximum (FWHM) of XRD peak with residual stress on ground surfaces. Philosophical Magazine 2012; 92(33): 4194-204.
International Atomic Energy Agency. Quality assurance programme for digital mammography. IAEA human health series no. 17. Vienna: IAEA; 2011. p. 3-24
Downloads
Published
How to Cite
Issue
Section
License
Copyright (c) 2023 BULLETIN OF THE DEPARTMENT OF MEDICAL SCIENCES

This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.