Comparison of Radiation Doses to Patients Pndergoing Standard Chest Radiographic Examinations with Conventional Screen-film Radiography and Computed Radiography
Main Article Content
Abstract
Background: New computed radiography equipment has recently been installed In Radiology Department of Buriram Hospital, its characteristics and versatility are well suited to the work undertaken in this environment.
Objective: The aim of this study was to compare radiation dose to patients undergoing standard chest radiographic examinations using conventional screen-film radiography and computed radiography.
Method: A observation study was analysed monthly between 1st October 2011 and 31st December 2011, entrance surface dose and effective dose were calculated for chest radiographic examinations using standard patient exposure parameters for the two imaging modalities.
Results: It was found that radiation doses for conventional screen-film radiography is higher than the doses for computed radiography, the radiation doses for conventional screen-film radiography and computed radiography were 0.15 mGy and 0.06 mGy, respectively. The image quality met the criteria in the European guidelines for all modalities.
Conclusions: During with time computed radiography systems was introduced to replace film-screen systems, its has been shown to have many advantages. The study compared radiation doses to patients undergoing standard chest radio graphic examinations using conventional screen-film radiography and computed radiography. Radiation doses for computed radiography was found 0.06 mGy lower than the doses for conventional screen-film radiography with acceptable image quality.
Article Details
References
2. Rowlands JA. The physics of computed radiography. Rhys Med Biol 2002; 47:123-66.
3. ชวลิต วงษ์เอก. เทคนิคการให้ปริมาณรังสีเพื่อการถ่ายภาพเอกซเรย์. [ออนไลน์]. [26 มิถุนายน 2556]; เข้าถึงได้จาก :URL: https://www.mt.mahidol.ac.th/e-learning/MTRD310/w%20first/f1.htm
4. Commission of the European Communities. CEC Quality Criteria for Diagnostic Radiographic Images and Patient Exposure Working Document, 3rd ed. CEC, Brussels, 1995
5. สุรวุฒิ บุญประกอบ. การเปรียบเทียบผลการได้รับปริมาณรังสีระหว่าง High kVp technique และ Low kVp technique ในผู้ป่วยถ่ายภาพรังสีทรวงอก. กลุ่มงานรังสีวิทยา โรงพยาบาลสิงห์บุรี. สิงห์บุรี;2550.
6. European Commission. Radiation Protection 109 : Guidance on Diagnostic Reference Levels for Medical Exposures. Belgium : European Communities, Luxembourg ; 1999.
7. ศิริวรรณ จูเลียง, ขวัญชัย วรากรศิริ. ปริมาณรังสีที่ผิวผู้ป่วยจากการถ่ายภาพรังสีวินิจฉัย ในโรงพยาบาลเขตชายฝั่งอันดามัน. วารสารวิชาการสาธารณสุข 2547;13:854-62.
8. Ng KH, Rassiah P, Wang HB, Hamlali AS, Muthuvellu P, Lee HP. Dose to patients in routine X-ray examinations in Malaysia. Br J Radiol 1998;71:654-60.
9. ศุภวิทู สุขเพ็ง, ธีราพร ตองติรัมย์, ประเวช แซงคำ, ไอริศรา ศิริสุนทร. ปริมาณรังสีดูดกลืนที่ผิวผู้ป่วย ได้รับต่อการถ่ายภาพเอกซเรย์ปอดของโรงพยาบาลในจังหวัดพิษณุโลก. วารสารวิชาการสาธารณสุข 2551; 17:59-67.